The device evaluation is anticipated.However, the complaint cannot not be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming with the evaluation when received.Lot number was not provided, therefore review of the manufacturing records could not be completed.The 510k number is not available as this is a japan custom defined product.
|
We received one single dpt kit with an iv set and pressure tubing for examination.The reported event of ¿incorrect flow rate¿ was not confirmed.As received into the laboratory, the snap tab was seated where it was supposed to be in the dpt housing.The snap tab functioned properly, and it was able to prime throughout the kit without any problem.It was also able to restrict the fast flush flow properly when the snap tab was not pulled during priming.No leakage was observed from the kit during leak test.All connections were tight and secure.The flow rate testing indicated that it was within specification.The reported event could not be confirmed or replicated during the analysis, as the device responded appropriately during functional testing.It is not known if some procedural factors may have contributed to the event.There was no evidence of a manufacturing nonconformance.No further actions will be taken at this time.With pressure tubing sets, it is common for the clinician to check for air or particulates while priming the line for use.Additionally, these products are used by highly trained clinicians, experienced in identifying and mitigating any hazards that arise.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised to consider the potential benefits in relation to the possible complications.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|