Background.Walker platform attachment issues reported by pts at (b)(6).Summary.Upon testing, the platform attachment is bending to the point that the metal attachment may split, making them unfit for patient use.During testing, the attachments were tightly secured and tested with weights well below the 300 lb capacity.Testing reports provided by pts at (b)(6).Testing also completed by (b)(6) pts.(b)(6) currently does not have walker platform attachments to provide to patient upon discharge as the equipment was deemed unfit for patients by their pts.Pts are providing platform attachments from the hospital or will request a home care coordinator to deliver a walker to the patient's home from another vendor.No patient harm as the patients have been discharged with the equipment.One incident report case entered.Scope: reports from pts at (b)(6) as well as (b)(6) hospital.Investigation: a platform walker attachment should carry 300 lbs.The platform should be bent 90 degrees and parallel to the ground.After a (b)(6) lb pt from (b)(6) tested the attachment, the platform bent to below 90 degrees.Identifiers on defective wheelchair saved by (b)(6) pt: drive manufacturer response (drive business manager): concern reported by other drive customers and, upon investigation, the attachments were not installed correctly, specifically regarding how tightly the equipment is screwed together.Testing by (b)(6) pts: pts found that, although the screws were tight and fully secure, the metal seemed to be flimsy and could easily snap with further weight applied to the platform.Patients who use the platform attachments will place all their weight on the attachments making the drive platforms unfit for patient use.Conclusion and next steps: drive walker platform attachments are not appropriate for patient use and should be pulled from inventory.Director of risk management to identify next steps in terms of reporting to (b)(4) or fda.Need long-term corrective action regarding inventory.
|