• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY FLEXTOME¿ CUTTING BALLOON¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY FLEXTOME¿ CUTTING BALLOON¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H749RB4350100
Device Problem Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/20/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was returned for analysis.The tip section of the device was visually and microscopically examined and no issues were noted that could have potentially contributed to the complaint incident.A visual and microscopic examination identified no issues with the blades.All blades were present and fully bonded to the balloon material.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood was identified within the balloon and lumen which is evidence of a device leak.The returned device was attached to an encore inflation unit and positive pressure was applied in an attempt to inflate the balloon.The balloon could not be inflated due to the presence of solidified blood that was present within the balloon and inflation lumen.The device was soaked in a water bath at a temperature of 37 degrees celsius to help soften the blood before further inflation attempts were made.The device was removed from the bath and the balloon was again attached to an inflation device subjected to positive pressure and liquid was observed to be leaking from a balloon pinhole approximately 7mm proximal to the proximal end of the distal marker band.An examination of the balloon material identified no issues which could potentially have contributed to this complaint.A visual and microscopic examination found no issues with the markerbands of the device.A visual and tactile examination of the device identified no kinks or damage along the shaft of the device.No other issues were identified during the product analysis.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event Description
Reportable based on device analysis completed on 07-dec-2017.It was reported that the balloon catheter failed to cross the lesion.The target lesion was located in the severely calcified artery.A 10/3.50 flextome¿ cutting balloon¿ was selected for use.During the procedure, it was noted that the balloon was unable to cross the target lesion due to severe calcification.The procedure was completed with a different device.No patient complications were reported.However, device analysis revealed a balloon pinhole approximately 7mm proximal to the proximal end of the distal marker band.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXTOME¿ CUTTING BALLOON¿
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7137858
MDR Text Key95666118
Report Number2134265-2017-12574
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
PMA/PMN Number
P950020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Remedial Action Inspection
Type of Report Initial
Report Date 12/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/21/2019
Device Model NumberH749RB4350100
Device Catalogue NumberRB435010
Device Lot Number0019505569
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2017
Is the Reporter a Health Professional? Yes