• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG QUART ARTERIAL FILTER WITH SOFTLINE COATING; FILTER, BLOOD, CARDIOPULMONARY BYPASS, ARTERIAL LINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG QUART ARTERIAL FILTER WITH SOFTLINE COATING; FILTER, BLOOD, CARDIOPULMONARY BYPASS, ARTERIAL LINE Back to Search Results
Model Number HBF 140-J
Device Problem Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/02/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Maquet getinge group is aware of similar complaints from this product.Similar products, showing a similar malfunction, have been tested.We tested the returned filter in our laboratory, following our standard process, on tightness.Therefore, the quart was tested at below water in a water bath and was then pressurized with compressed air (0.3 bar).Due to the air escaping (visible as air bubbles in water) a leakage in the welding cover has been found.The leakage at the connection between cover and filter body can be confirmed.Most possible root cause could be the bad welding between cover and filter body.Since the reported failure did not contribute to a death or serious injury no corrective action is needed.In addition at this time it cannot be concluded that this is an systemic error.This data will be handled through a designated maquet trending process.If a trend occurs, it will be escalated to quality assurance management for review and determination if further investigation is necessary.
 
Event Description
Customer reported that "during priming at (b)(6), it was noted that the priming solution was leaking from the upper side of the quart." no known consequences to the patient.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUART ARTERIAL FILTER WITH SOFTLINE COATING
Type of Device
FILTER, BLOOD, CARDIOPULMONARY BYPASS, ARTERIAL LINE
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key7138656
MDR Text Key95791690
Report Number8010762-2017-00406
Device Sequence Number1
Product Code DTM
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K090518
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/01/2018
Device Model NumberHBF 140-J
Device Catalogue Number701031617
Device Lot Number70112831
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/21/2017
Initial Date FDA Received12/21/2017
Date Device Manufactured06/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-