• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG. MONOPOLAR ELECTROSURGICAL PROBE, L-TIP, 33CM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG. MONOPOLAR ELECTROSURGICAL PROBE, L-TIP, 33CM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 0250040112
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Information (3190)
Event Date 11/30/2017
Event Type  malfunction  
Manufacturer Narrative
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.(b)(4).
 
Event Description
It was reported that the device failed the insulation test.
 
Manufacturer Narrative
The product was returned for investigation and the reported failure mode was confirmed.The reported failure mode will be monitored for future reoccurrence.Alleged failure: these failed the insulation test way too early.The failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be user misuse, excessive force, normal wear, or improper sterilization methods.Mfg date: manufacture date is not known.(b)(4).
 
Event Description
It was reported that the device failed the insulation test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG. MONOPOLAR ELECTROSURGICAL PROBE, L-TIP, 33CM
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
roomi banerjee dua
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key7139301
MDR Text Key95834346
Report Number0002936485-2017-01234
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K934094
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0250040112
Device Lot Number272621
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/30/2017
Initial Date FDA Received12/21/2017
Supplement Dates Manufacturer Received11/30/2017
Supplement Dates FDA Received03/23/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-