During the procedure, a shift on the left pulmonary veins occurred after approximately one hour.Everything appeared as expected after creating the geometry of the left atrium and the right veins were isolated.After the catheter was moved to the left side veins, the catheter appeared to be approximately 1cm outside of both veins, which was inaccurate.A new geometry was obtained and the procedure was completed.The procedure was extended by 30 minutes due to this issue.
|
The review of the electronic files confirmed that catheter display shift occurred during the clinical case; the event was verified.The system continues to remain in use at the account.The following are potential reasons why a catheter shift may occur clinically( this list is not inclusive): patient repositioning such as upper limb motion and displacement of the chest or abdomen.Positional reference patch dislodgement.Non-transient shifts can be related to the gravitational shift of body weight as the patient relaxes during the procedure and subsequent gradual impedance changes in the body.Patch connection, adhesion, or placement issues.Powering up or shutting down other lab equipment can cause electro-magnetic interference with the abbott equipment.In addition, changing connections to other lab equipment such as a stimulator or generator can cause catheter shift.
|