Model Number 60-01-04
Device Problem No Flow (2991)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/27/2017
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.Livanova (b)(4) manufactures the centrifugal pump 5 (cp5) the incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.Visual inspection did not identify any abnormalities or defects and several hours of troubleshooting was unable to duplicate the reported issue.The user was not comfortable returning the pump to service, so it was replaced and is planned to be returned to livanova for further investigation.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Evaluated on site by livanova technician.
 
Event Description
Livanova (b)(4) received a report that the centrifugal pump 5 (cp5) displayed a pressure error and forward flow stopped during a case, resulting in a negative flow alarm.While troubleshooting the system circuit and other parts of the oxygenator, the centrifugal pump began forward flow again.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP 5 (CP5)
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key7139619
MDR Text Key95872382
Report Number9611109-2017-01040
Device Sequence Number1
Product Code DWA
UDI-Device Identifier04033817900894
UDI-Public010403381790089411151103
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 12/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-01-04
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/27/2017
Initial Date FDA Received12/21/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/28/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-