• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY FLEXTOME¿ CUTTING BALLOON¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY FLEXTOME¿ CUTTING BALLOON¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H749RB4300100
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/13/2017
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4).The device was not be returned for evaluation.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event Description
It was reported that a balloon rupture occurred.The concentric, de novo target lesion was located in the mildly calcified coronary artery.A 10/3.00 flextome¿ cutting balloon¿ was selected for use.During the procedure, the balloon burst when it was inflated.The procedure was completed with another of the same device.No patient complications were reported and patient status is stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXTOME¿ CUTTING BALLOON¿
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7139810
MDR Text Key95664002
Report Number2134265-2017-13008
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
P950020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Physician
Type of Report Initial
Report Date 12/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/21/2020
Device Model NumberH749RB4300100
Device Catalogue NumberRB430010
Device Lot Number0020671936
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/14/2017
Initial Date FDA Received12/21/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/15/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-