(b)(4).Returned product consisted of a ffr comet pressure wire.The tip, device shaft, sensor port and the coefficient values was examined for damage or any irregularities.The guidewire shaft showed 1 kink.The kink was located 47cm from the tip.It was also noticed that some of the coating on the wire was peeled in the same location of the kink.No other damage was noticed on the shaft.The sensor port was clear of any residue.The pressure wire was connected to the analysis support test bench and all applicable data was correct as designed, there was no difficulty in connecting the wire.The coefficient was confirmed to be in specification.The investigation conclusion is handling damage as the complaint was caused by handling of the device or portion of the device without direct patient contact.(b)(4).
|
Reportable based on analysis completed on 14-dec-2017.It was reported that the shaft of the wire had a bend.During introduction of a comet 185cm pressure guidewire, a bend in the middle of the shaft was noted which prevented the wire from being pushed through the tuohy at the bend location.The device was removed and the procedure completed with another of the same device.No patient complications were reported and the patient's status was fine.However, returned device analysis revealed peeled coating.
|