• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. SMARTSTITCH PP CONNECTOR; ACCESSORIES,ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. SMARTSTITCH PP CONNECTOR; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Catalog Number OM-8010
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
It was reported that the top jaw of the device broke off in the patient when it was removed from the cannula.The broken piece was removed.No patient injury or further complications reported.
 
Manufacturer Narrative
The reported smartstitch pp connector, intended for use in treatment, was not returned for evaluation.A relationship between the product and reported incident cannot be established as the product was not returned.Without the reported product a visual and functional evaluation cannot be performed and customers complaint cannot be confirmed.From the information provided, the top jaw of the device broke off in the patient when it was removed from the cannula.The broken piece was removed.An exact root cause cannot be determined without evaluation of the device; however, factors unrelated to the manufacture or design of the device that could have contributed to the reported event includes: excessive force.Excessive force can result in damage to the device.The instruction for use (ifu) were reviewed and were found to outline precautionary statements and instructions in regards to the use of the device to avoid damage or non-functionality.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTSTITCH PP CONNECTOR
Type of Device
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
jim gonzales
7000 w. william cannon
austin, TX 78735
MDR Report Key7141468
MDR Text Key95850887
Report Number3006524618-2017-00437
Device Sequence Number1
Product Code NBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123268
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 01/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberOM-8010
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/05/2017
Initial Date FDA Received12/22/2017
Supplement Dates Manufacturer Received01/19/2018
Supplement Dates FDA Received01/29/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-