|
Model Number H749236310040 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/08/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Age at time of event: 18 years or older.(b)(4).
|
|
Event Description
|
Same case as mdr id: 2134265-2017-13098.It was reported that the tip of the rotawire was rotating inside the blood vessel.The target lesion was located in the severely calcified left main trunk.A 1.75mm rotalink¿ plus and a 330cm rotawire¿ were selected for use.During the procedure, a 0.014" non bsc guidewire was advanced to the target lesion and then a non bsc microcatheter was inserted and replaced with the rotawire.When the rotawire was advanced inside the guiding catheter in order to reach the 1.75mm rotaburr up to lesion by dynaglide, it was noted under fluoroscopy that the tip of the rotawire was found rotating in the blood vessel.Afterwards, the rotawire was slightly withdrawn and was advanced again through dynaglide mode.Ablation was then performed and was completed with the same rotalink and the same rotawire.No patient complications were reported.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer: the device was returned for analysis.Returned product consisted of the rotablator rotalink plus device.The brake button, advancer, handshake connections, sheath, coil and burr were microscopically and visually inspected.Microscopic examination of the device revealed that the annulus was damaged and not rounded.The damage to the annulus is consistent with damage caused by contact with the rotawire during use.The brake button was received in the as-manufactured position.The brake button was tested and it was able to be pressed and released with no issues.The guidewire used in the procedure was not returned for product analysis, so functional testing was completed with a test rotawire.The rotawire was not able to be inserted in the annulus of the burr due to the damage, so the advancer and burr catheter were detached and the rotawire was inserted into the handshake connection of the advancer.Functional testing was performed by connecting the advancer device to the rotablator control console system and the brake was tested with the rotawire.The brake and rotawire worked properly with no issues.The advancer was opened and the brake was examined.There was no damage or irregularities to the brake or the inside the advancer at the brake location.Inspection of the remainder of the device presented no other damage or irregularities.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
|
|
Event Description
|
Same case as mdr id: 2134265-2017-13098.It was reported that the tip of the rotawire was rotating inside the blood vessel.The target lesion was located in the severely calcified left main trunk.A 1.75mm rotalink¿ plus and a 330cm rotawire¿ were selected for use.During the procedure, a 0.014" non bsc guidewire was advanced to the target lesion and then a non bsc microcatheter was inserted and replaced with the rotawire.When the rotawire was advanced inside the guiding catheter in order to reach the 1.75mm rotaburr up to lesion by dynaglide, it was noted under fluoroscopy that the tip of the rotawire was found rotating in the blood vessel.Afterwards, the rotawire was slightly withdrawn and was advanced again through dynaglide mode.Ablation was then performed and was completed with the same rotalink and the same rotawire.No patient complications were reported.
|
|
Search Alerts/Recalls
|
|
|