TERUMO CARDIOVASCULAR SYSTEMS CORPORATION LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP
|
Back to Search Results |
|
Model Number 801041 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/06/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Evaluation is in progress, but not yet concluded.The fsr removed the suspect pump and returned it to the manufacturer for further evaluation.The unit operated to the manufacturer's specifications.
|
|
Event Description
|
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, there was a pump jam error message with no tubing in the raceway.There was no patient involvement.
|
|
Manufacturer Narrative
|
(b)(4).The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed pin#4 of the optical sensor cable to be disengaged from connector p11, which resulted in pump jam.Reinsertion of pin#4 into connector p11 enabled pump to function properly.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|