• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DORNOCH ULTRAFLEX DUO HIGH FLUID CART WITH SMOKE EVACUATOR; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DORNOCH ULTRAFLEX DUO HIGH FLUID CART WITH SMOKE EVACUATOR; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Model Number N/A
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/04/2017
Event Type  malfunction  
Manufacturer Narrative
This complaint is recorded with zimmer biomet under (b)(4).Pma 510(k): k081047; k123188; k133786.The product will be evaluated by an external contractor.Once the evaluation/investigation has been completed, a follow up mdr will be submitted.Product evaluated by external contractor.
 
Event Description
It was reported that the unit was leaking from bottom of the cart.No adverse events have been reported as a result of this malfunction.
 
Manufacturer Narrative
This complaint is recorded with zimmer biomet under (b)(4).The previous repair report for ultra duo flex fluid cart serial number (b)(4) was reviewed and noted no related non-conformances, requests for deviation (rfd) or any other issues with the repair.The previous repair report review found no issues with the device after repair and all verifications, inspections and tests were successfully completed.Using crm to query for serial number (b)(4), the device was noted to have been previously repaired three times, the previous repair being for an issue with a valve pack on 16 november 2017.The valve pack is not associated with the current repair.Thus, this repair was a non-related issue.On (b)(6) 2017, it was reported from university of (b)(6) that a unit was leaking from bottom of cart.Day mechanical systems, inc was contacted about the cart and dispatched a service technician to be at the site.The technician arrived at the site and confirmed that the level sensor was leaking.He replaced the level sensor (part # 91584) and the sound insulation (part # 70011) and then verified that the cart was functioning as intended.The technician then returned the cart to service without further incident.The device was tested, inspected, and repaired as per cl ¿ repair cart and evac rev.0.The root cause for the unit leaking from bottom of the cart was due to a loose level sensor.The reported event was confirmed during inspection of the device and the device was noted to be functioning as intended after the level sensor was replaced.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX DUO HIGH FLUID CART WITH SMOKE EVACUATOR
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer (Section G)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key7141877
MDR Text Key95954388
Report Number0001954182-2017-00045
Device Sequence Number1
Product Code JCX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PSEE H10
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberULDU500SE
Device Lot Number0019210
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/04/2017
Initial Date FDA Received12/22/2017
Supplement Dates Manufacturer Received01/05/2018
Supplement Dates FDA Received01/09/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/14/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-