ST. JUDE MEDICAL, INC. WORKMATE¿ CLARIS¿ DISPLAY PLUS AMPLIFIER; PROGRAMMABLE DIAGNOSTIC COMPUTER Back to Search Results
Model Number H700150
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/12/2017
Event Type  Injury  
Event Description
During device preparation, with the patient prepped under general anesthesia, an amplifier error message displayed that read, "amplifier is off or not connected, please check the connection".The device was power cycled but the issue remained, so the procedure was cancelled.There were no adverse consequences to the patient.
 
Manufacturer Narrative
Based on the results of the hardware evaluation performed, it was concluded that the returned amplifier functioned as expected.The hardware evaluation focused on the basic functions of the returned amplifier such as communication, signals acquisition/amplification and data processing.No abnormal operating symptoms or unexpected abnormal functions under normal operating condition were observed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with abbott specifications and procedures.Based on the information provided and the hardware evaluation performed, the cause of the reported communication issue could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WORKMATE¿ CLARIS¿ DISPLAY PLUS AMPLIFIER
Type of Device
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key7146259
MDR Text Key95768226
Report Number2184149-2017-00079
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132073
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberH700150
Device Catalogue NumberH700150
Device Lot Number5544879
Other Device ID Number05415067001504
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/12/2017
Initial Date FDA Received12/26/2017
Supplement Dates Manufacturer Received01/08/2018
Supplement Dates FDA Received01/25/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/29/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-