Catalog Number DP-48K
Device Problem Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/14/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device history record of batch number provided has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No rejection report was originated for the lot in question that can be associated to the complaint reported.The dhr shows that the product was assembled and inspected according to our specifications.The device has not been returned for investigation at this time.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that when the aortic punch was used, the punch could be released once, then it jammed and could not be released even with difficulty.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PU DP-48K DISP PUNCH 4.8MM
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
effie jefferson
3015 carrington mill blvd
morrisville 27560
9194332672
MDR Report Key7147849
MDR Text Key95962395
Report Number3004365956-2017-00463
Device Sequence Number1
Product Code DWS
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/07/2020
Device Catalogue NumberDP-48K
Device Lot Number74F1503522
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/15/2017
Initial Date FDA Received12/27/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/29/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-