Catalog Number 213000000
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2017
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Surgeon adjusting trial cylinder during fitting and the spring separated from the device.He also noted that it initially was not snapping into the trial stem as per usual.Minimal delay did not affect surgery as another device was available.No issues with the new trial assembly.Patient consequence? :no.Is the information being submitted for this complaint all the details that are known/available regarding this event? : yes.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: examination of the returned device confirmed the reported event.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLOBAL AP BALL CYLIN TRL ASSEM
Type of Device
EXTREMITY INSTRUMENTS : HUMERAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581-0988
5743725905
MDR Report Key7150538
MDR Text Key95931296
Report Number1818910-2017-52579
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295090366
UDI-Public10603295090366
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 11/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number213000000
Device Lot NumberHG4666
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/29/2017
Initial Date FDA Received12/28/2017
Supplement Dates Manufacturer Received01/30/2018
08/02/2018
Supplement Dates FDA Received01/31/2018
08/03/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/18/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-