COOK ENDOSCOPY ACROBAT¿ CALIBRATED TIP WIRE GUIDE; OCY, ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Catalog Number ACRO-35-450
Device Problem Peeled/Delaminated (1454)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/04/2017
Event Type  malfunction  
Manufacturer Narrative
Concomitant products: boston scientific cutting wire.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.If additional pressure is applied to the wire guide and/or accessory device(s) while moving the wire guide inside the accessory device(s), this could contribute to wire guide coating damage.The instructions for use instruct the user to do the following: "prior to removing wire guide from holder, flush with 30 cc of sterile water." failure to flush the wire guide can result in damage to the wire guide."flush endoscope accessory channel and/or lumen of device with sterile water, then insert wire guide floppy end first.Note: for best results, wire guide should be kept wet, if applicable." failure to flush the endoscope channel can result in damage to the wire guide.Prior to distribution, all acrobat calibrated tip wire guides are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook acrobat calibrated tip wire guide.The wire guide covering peeled and detached 27 cm from the head of wire guide.
 
Manufacturer Narrative
Continued from concomitant products: boston scientific cutting wire investigation evaluation: our evaluation of the product said to be involved confirmed the report.Approximately 26.5 cm to 26.6 cm from the distal end, the wire guide covering has folded over itself.Approximately 26.6 cm to 28.0 cm from the distal end is a section of bare core wire.A section of the coating approximately 1.1 cm long is hanging from the wire guide, the coating is still attached at approximately 28.0 cm from the distal end.Due to the condition of the returned device it cannot be determined if any sections of the coating are missing.The wire guide covering feels rough approximately 28.0 cm to 28.9 cm and 221 cm to 225 cm from the distal end and 81.0 cm to 86.0 cm from the proximal end.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.If additional pressure is applied to the wire guide and/or accessory device(s) while moving the wire guide inside the accessory device(s), this could contribute to wire guide coating damage.The instructions for use instruct the user to do the following: "prior to removing wire guide from holder, flush with 30 cc of sterile water." failure to flush the wire guide can result in damage to the wire guide."flush endoscope accessory channel and/or lumen of device with sterile water, then insert wire guide floppy end first.Note: for best results, wire guide should be kept wet, if applicable." failure to flush the endoscope channel can result in damage to the wire guide.Prior to distribution, all acrobat calibrated tip wire guides are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACROBAT¿ CALIBRATED TIP WIRE GUIDE
Type of Device
OCY, ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key7151400
MDR Text Key96396937
Report Number1037905-2017-00790
Device Sequence Number1
Product Code OCY
UDI-Device Identifier00827002342668
UDI-Public(01)00827002342668(17)200602(10)W3870347
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 01/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberACRO-35-450
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date12/04/2017
Device Age6 MO
Event Location Hospital
Initial Date Manufacturer Received 12/05/2017
Initial Date FDA Received12/28/2017
Supplement Dates Manufacturer Received01/03/2018
Supplement Dates FDA Received01/25/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
OLYMPUS JF-260V ENDOSCOPE
Patient Age70 YR
Patient Weight52
-
-