Model Number 3100A
Device Problems Device Alarm System (1012); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 12/01/2017
Event Type  malfunction  
Manufacturer Narrative
Vyaire medical complaint number (b)(4).At this time, vyaire medical has not received the suspect device/component from the customer for evaluation.In the event that the device is received for evaluation or additional information is received, a follow-up report will be submitted.
 
Event Description
The customer reported that there was no alarm and the silence button latched for more than 45 seconds.The customer pressed the start/stop button and the ventilator started to alarm.When the customer pressed the alarm silence button to silence, the alarm silenced for 45 seconds as per specifications.There was no patient involvement associated with this issue.
 
Manufacturer Narrative
A vyaire medical field service representative (fsr) evaluated the device onsite.The fsr was unable to duplicate the reported issue as the original settings on the ventilator at the time of the reported issue were not preserved.The fsr adjusted the pressure transducer and zero and span pot.The fsr performed the patient circuit calibration and performance check.The device meets manufacturer's specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATORY VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
VYAIRE MEDICAL, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
kristin graf
22745 savi ranch parkway
yorba linda, CA 92887
MDR Report Key7151678
MDR Text Key95955671
Report Number2021710-2017-07226
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/01/2017
Initial Date FDA Received12/28/2017
Supplement Dates Manufacturer Received12/28/2017
Supplement Dates FDA Received01/24/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-