MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 3058
Device Problems Failure to Deliver Energy (1211); Device Operates Differently Than Expected (2913)
Patient Problems Incontinence (1928); Therapeutic Effects, Unexpected (2099)
Event Type  Injury  
Manufacturer Narrative
Information references the main component of the system and other applicable components are: product id: 3889-28, lot# va05yv3, implanted: (b)(6) 2013, product type: lead.The codes listed apply to both lead and ins.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer via a manufacturer¿s representative regarding a patient with an implanted neurostimulator ( ins) for gastrointestinal /pelvic floor therapy.It was reported that the patient had been leaking for a while and wore pads.When their battery and lead were replaced it was discovered that the lead wire was "misplaced", meaning it was totally out of place, which is the reason it was not working and the patient wasn¿t getting the stimulation she needed.The healthcare professional (hcp) discovered the problem with the lead during a fluoroscopy and said the lead was pointing in a totally opposite direction than it should be for the nerve to be affected.Caller states the hcp said the lead was not broken but just out of place but the doctor opted to put in a new lead and ins.No further complications were reported or are anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM II
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7152341
MDR Text Key96003332
Report Number3004209178-2017-26759
Device Sequence Number1
Product Code EZW
UDI-Device Identifier00613994913654
UDI-Public00613994913654
Combination Product (y/n)N
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 12/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/14/2014
Device Model Number3058
Device Catalogue Number3058
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/27/2017
Initial Date FDA Received12/28/2017
Date Device Manufactured01/22/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-