MEDTRONIC MED REL MEDTRONIC PUERTO RICO INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
|
Back to Search Results |
|
Model Number 3058 |
Device Problems
Failure to Deliver Energy (1211); Device Operates Differently Than Expected (2913)
|
Patient Problems
Incontinence (1928); Therapeutic Effects, Unexpected (2099)
|
Event Type
Injury
|
Manufacturer Narrative
|
Information references the main component of the system and other applicable components are: product id: 3889-28, lot# va05yv3, implanted: (b)(6) 2013, product type: lead.The codes listed apply to both lead and ins.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a consumer via a manufacturer¿s representative regarding a patient with an implanted neurostimulator ( ins) for gastrointestinal /pelvic floor therapy.It was reported that the patient had been leaking for a while and wore pads.When their battery and lead were replaced it was discovered that the lead wire was "misplaced", meaning it was totally out of place, which is the reason it was not working and the patient wasn¿t getting the stimulation she needed.The healthcare professional (hcp) discovered the problem with the lead during a fluoroscopy and said the lead was pointing in a totally opposite direction than it should be for the nerve to be affected.Caller states the hcp said the lead was not broken but just out of place but the doctor opted to put in a new lead and ins.No further complications were reported or are anticipated.
|
|