ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FORTIFY ASSURA DR ICD, US; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD2357-40Q
Device Problems Premature Discharge of Battery (1057); Failure to Interrogate (1332)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/19/2017
Event Type  Injury  
Event Description
It was reported that the patient presented in the clinic and the implantable cardiac defibrillator could not be interrogated.Premature battery depletion was suspected.The device was explanted and replaced.The patient was stable prior, during and post procedure.
 
Manufacturer Narrative
Premature battery depletion was confirmed by analysis.No sources of high current were noted.The cause of the premature battery depletion was consistent with li cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY ASSURA DR ICD, US
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7153456
MDR Text Key96001242
Report Number2938836-2017-35485
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 08/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2017
Device Model NumberCD2357-40Q
Device Lot NumberA000003065
Other Device ID Number05414734508094
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/19/2017
Initial Date FDA Received12/28/2017
Supplement Dates Manufacturer Received07/12/2018
Supplement Dates FDA Received08/03/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/12/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0115-2017
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age71 YR
-
-