• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number X SERIES
Device Problem Failure of Device to Self-Test (2937)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event Description
Complainant alleged that during a routine shift check by a clinician, the device failed self-test for ecg function.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Manufacturer Narrative
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a faulty integrated circuit on the monitor board.Analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key7153834
MDR Text Key96009755
Report Number1220908-2017-03307
Device Sequence Number1
Product Code DRT
UDI-Device Identifier00847946014865
UDI-Public00847946014865
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112432
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/08/2017
Initial Date FDA Received12/29/2017
Supplement Dates Manufacturer Received12/08/2017
Supplement Dates FDA Received01/17/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-