Results: bd received one sample for investigation.Samples were decontaminated.A review of the device history record revealed no irregularities during the manufacture of the reported lot # 7017002.The returned material showed the reported defect.Upon visual examination, the extend tubing was found broken in the middle.The deformation looks to be caused by the slide clamp.The od and id thickness was measured and was found in uniformity of the extend tubing.Microscopic examination revealed no abnormality when measuring the part line flash dimension of the slide clamp.The defect may have been caused by the slide clamp design, the angle of the slide clamp or the material of the extend tubing.However, no abnormalities were found in any of the parameters of the returned sample.Therefore, a root cause cannot be finalized.
|
It was reported that when a nurse performed a venipuncture on a patient using a bd pegasus¿ safety closed iv catheter system, the nurse found blood spilled as she opened the clip.As she pulled the device out, she found a hole on the tubing near the hub.There was no report of injury or medical intervention.
|