Catalog Number 2C6750H
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/01/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported an air in line alarm occurred on a sigma spectrum pump while using an interlink blood recipient set.This event occurred during patient infusion.Blood was dripping freely, but when placed on the iv pump, it would constantly beep and say ¿air in line¿.There was no visible issue with the tubing that they could see, there was no air in line, the pump appeared to be clean and in good working order.The issue was resolved after they had changed the tubing.The pump would beep with the air in line alarm within a minute of starting the infusion.All connections were checked to ensure they were secure prior to infusion.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not returned; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLINK BLOOD RECIPIENT SET
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
haina san cristobal
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina san cristobal
DR  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7155431
MDR Text Key96068743
Report Number1416980-2017-10074
Device Sequence Number1
Product Code BRZ
UDI-Device Identifier00085412073583
UDI-Public(01)00085412073583
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K993120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date09/23/2022
Device Catalogue Number2C6750H
Device Lot NumberDR17I22014
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/05/2017
Initial Date FDA Received12/29/2017
Supplement Dates Manufacturer Received01/11/2018
Supplement Dates FDA Received01/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/22/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
SIGMA SPECTRUM PUMP
-
-