Model Number 302-20 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Muscle Spasm(s) (1966); Neck Pain (2433); Fibrosis (3167); No Code Available (3191)
|
Event Date 04/01/2015 |
Event Type
Injury
|
Event Description
|
It was reported that the patient had issues with a lead pulling sensation.The patient had a revision surgery to loosen up the lead for more range of motion for the patient and it was noted that the patient had fibrosis around the lead.No additional relevant information has been received to date.
|
|
Manufacturer Narrative
|
Describe event or problem - corrected information: the information included in describe event or problem on this report should have been included in the initial report as the surgery for the patient was due to the all of the adverse events reported for the patient.Relevant tests/laboratory data - corrected information: the settings and diagnostic data should have been included in the initial report.(b)(4).
|
|
Event Description
|
Information should have been included in the initial report that the patient also had painful stimulation and muscle spasms in the neck at the electrode site.The spasms in the patient's neck subsided when the vns magnet was placed over the generator to disable it.The patient's settings were adjusted multiple times and the normal and systems diagnostics were stated to be ok.The painful stimulation and muscle spasms were stated to not have resolved after the surgery to reposition the lead indicating that the surgery could have been to help alleviate those adverse events as well.
|
|