BOSTON SCIENTIFIC - MARLBOROUGH UPHOLD¿ LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN
|
Back to Search Results |
|
Model Number M0068318170 |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/12/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Analysis of the returned device revealed that the suture on the blue with white stripe dilator was broken.The dart was retrieved from behind the cage of the capio slim.The dart has a small amount of suture attached to it.Analysis also revealed that the dart on the blue dilator was caught as intended in the capio cage when actuated and there was no damage to the capio slim suture capturing device.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.The investigation concluded that this complaint is associated with a product that meets the design and manufacture specifications but due to anatomical/procedural factors encountered during the procedure, performance of the device was limited.The most probable root cause classification is operational context.
|
|
Event Description
|
It was reported to boston scientific corporation that an uphold¿ lite device was used during a sacrospinous ligament fixation procedure performed on (b)(6) 2017.According to the complainant, during procedure, the capio cage failed to catch the needle.The procedure was completed with another uphold¿ lite device.The patient's condition at the conclusion of the procedure was reported to be stable.This event has been deemed reportable based on the investigation results: the suture on the blue dilator is broken.The dart was retrieved from behind the cage of the capio slim.The dart has a small amount of suture attached to it.
|
|
Search Alerts/Recalls
|
|
|