Model Number VICMO12.1 |
Device Problem
Inadequacy of Device Shape and/or Size (1583)
|
Patient Problem
No Code Available (3191)
|
Event Date 11/10/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
The product is manufactured in the us, but not marketed in the us.(b)(4).Conclusion code: off-label use [under 21 years of age at time of implant ((b)(6)), anterior endothelium chamber depth < 3.0mm (2.91mm)].(b)(4).
|
|
Event Description
|
The reporter indicated the surgeon implanted a 12.1mm vicmo12.1 implantable collamer lens, diopter -6.0 into the patient's right eye (od) on (b)(6) 2017.On (b)(6) 2017, the lens was exchanged with a longer lens due to low vault.The problem was resolved.
|
|
Manufacturer Narrative
|
The product was returned dry, in a micro centrifuge vial with clear surgical residue/debris on product.Visual inspection found the haptic bent and residue on the lens surface.(b)(4).
|
|