• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VTICMO12.6
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Code Available (3191)
Event Date 09/05/2017
Event Type  Injury  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.(b)(4).Conclusion code: off-label use [anterior endothelium chamber depth < 3.0mm (2.89mm)].(b)(4).
 
Event Description
The reporter indicated that the surgeon implanted a 12.6mm vticmo12.6 implantable collamer lens, -17.0/+1.50/099 (sphere/cylinder/axis), into the patient's left eye (os) on (b)(6) 2017.On (b)(6) 2017 the lens was exchanged with a shorter lens due to excessive vault.The problem is resolved.
 
Manufacturer Narrative
Device evaluation: the lens was returned dry, in a micro centrifuge vial.There was residue/debris on product.Visual inspection found no visible damage to the lens and foreign material on lens surface (residue and debris on lens).(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key7156013
MDR Text Key96109395
Report Number2023826-2017-01969
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date08/31/2019
Device Model NumberVTICMO12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/06/2017
Initial Date FDA Received12/29/2017
Supplement Dates Manufacturer Received03/21/2018
Supplement Dates FDA Received04/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/20/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age25 YR
-
-