GENZYME BIOSURGERY (RIDGEFIELD) SYNVISC ONE; INTRA-ARTICULAR HYALURONIC ACID Back to Search Results
Lot Number 7RSL021
Device Problems Device Contamination With Biological Material (2908); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Arthralgia (2355); Joint Swelling (2356)
Event Date 11/06/2017
Event Type  malfunction  
Event Description
This case was cross referenced with (b)(4) (cluster).This unsolicited case from united states was received on (b)(6) 2017 from the health care professional.This case concerns an (b)(6) years old female patient who received treatment with synvisc one and later after few days had left knee pain and swelling.No relevant medical history, past medications, concurrent conditions and concomitant medications were reported.On (b)(6) 2017, the patient received treatment with intra-articular synvisc one injection, once at a dose of 6 ml (lot number: 7rsl021; expiration date: may 31, - 2020) for osteoarthritis into the left knee.On an unknown date in (b)(6) 2017, few days after starting treatment the patient had left knee pain and swelling.On (b)(6) 2017, it was reported that the patient was given methylprednisolone acetate (depo-medrol) injection 40 mg.Corrective treatment: methylprednisolone acetate injection for both the events.Outcome: unknown for both the injections seriousness criteria: required intervention for both the events a pharmaceutical technical complaint (ptc) was initiated with ptc number: (b)(4).
 
Event Description
This case was cross referenced with (b)(4).This unsolicited case from united states was received on 06-dec-2017 from the health care professional.This case concerns an (b)(6) years old female patient who received treatment with synvisc one and later after few days had left knee pain and swelling, also, device malfunction was identified for the reported lot number.No relevant medical history, past medications, concurrent conditions and concomitant medications were reported.On (b)(6) 2017, the patient received treatment with intra-articular synvisc one injection, once at a dose of 6 ml (lot number: 7rsl021; expiration date: 31-may- 2020) for osteoarthritis into the left knee.On an unknown date in (b)(6) 2017, few days after starting treatment the patient had left knee pain and swelling.On (b)(6) 2017, it was reported that the patient was given methylprednisolone acetate (depo-medrol) injection 40mg.Corrective treatment: not reported for device malfunction; methylprednisolone acetate injection for other events.Outcome: unknown for all events.Seriousness criteria: required intervention for all the events.A pharmaceutical technical complaint (ptc) was initiated with ptc number: (b)(4).An investigation was initiated as a result of an unexpected increase in the number of labelled adverse events received from the us market for synvisc one, lot 7rsl021.The product met all release testing at time of manufacture in june 2017.Retain samples were retested due to the unexpected increase in adverse events.Higher than expected endotoxin results were obtained.In addition, the presence of microbial contamination was also confirmed.The cause of these events is under investigation.Once this investigation is completed, corrective and preventive actions will be implemented.Additional information was received on 06-dec-2017 and 02-jan-2018 (processed with clock start date of 06-dec- 2017).Global ptc number and ptc results were added.An additional event of device malfunction was added.Pharmacovigilance comment: sanofi company comment for follow up dated 06-dec-2017: this case concern a female patient who received synvisc one injection from the recalled lot and had left knee swelling and pain.A temporal relationship can be established with the product administration.Furthermore, the concerned lot number has been identified to have malfunction by the company.Therefore, the causal relationship of the events to the products cannot be excluded.
 
Event Description
This case was cross referenced with (b)(4).This unsolicited case from united states was received on 06-dec-2017 from the health care professional.This case concerns an (b)(6) years old female patient who received treatment with synvisc one and on the same day had left knee pain and swelling, also, device malfunction was identified for the reported lot number.No relevant past medications were reported.Patient had no known drug allergies.Patient had carpal tunnel, carpal tunnel and osteoarthritis knee.Concomitant medications include methotrexate, infliximab (remicade), solifenacin succinate (vesicare), aledronate sodium and naproxen sodium (aleve).On (b)(6) 2017, at 08:53, the patient received treatment with intra-articular synvisc one injection, once at a dose of 6 ml (lot number: 7rsl021; expiration date: 31-may-2020) for osteoarthritis into the left knee.On the same day, after starting treatment the patient had left knee pain and swelling.It was reported that the patient had worsening knee pain and swelling.On (b)(6) 2017, it was reported that the patient was given methylprednisolone acetate (depo-medrol) injection 40mg.Corrective treatment: not reported for device malfunction; methylprednisolone acetate injection for other events outcome: recovered for all events seriousness criteria: required intervention for all the events a pharmaceutical technical complaint (ptc) was initiated with ptc number: (b)(4) an investigation was initiated as a result of an unexpected increase in the number of labelled adverse events received from the us market for synvisc one, lot 7rsl021.The product met all release testing at time of manufacture in june 2017.Retain samples were retested due to the unexpected increase in adverse events.Higher than expected endotoxin results were obtained.In addition, the presence of microbial contamination was also confirmed.The cause of these events is under investigation.Once this investigation is completed, corrective and preventive actions will be implemented.Additional information was received on 06-dec-2017 and 02-jan-2018 (processed with clock start date of 06-dec- 2017).Global ptc number and ptc results were added.An additional event of device malfunction was added.Additional information was received on 19-jan-2018.Worsening knee pain and swelling was added as symptom of left knee swelling.Event start dates was updated.Concomitant medication and medical history added.Clinical course updated.Text was amended accordingly.Pharmacovigilance comment: sanofi company comment for follow up dated 06-dec-2017: this case concern a female patient who received synvisc one injection from the recalled lot and had left knee swelling and pain.A temporal relationship can be established with the product administration.Furthermore, the concerned lot number has been identified to have malfunction by the company.Therefore, the causal relationship of the events to the products cannot be excluded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNVISC ONE
Type of Device
INTRA-ARTICULAR HYALURONIC ACID
Manufacturer (Section D)
GENZYME BIOSURGERY (RIDGEFIELD)
1125 pleadantview terrace
ridgefield NJ 07657
MDR Report Key7156218
MDR Text Key96112282
Report Number2246315-2017-00279
Device Sequence Number1
Product Code MOZ
Combination Product (y/n)N
PMA/PMN Number
P940015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 12/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date05/31/2020
Device Lot Number7RSL021
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/06/2017
Initial Date FDA Received12/30/2017
Supplement Dates Manufacturer Received12/06/2017
01/19/2018
Supplement Dates FDA Received01/12/2018
02/12/2018
Patient Sequence Number1
Treatment
ALEDRONATE SODIUM(CON.); ALEVE(CON.); METHOTREXATE(CON.); REMICADE(CON.); VESICARE(CON.)
Patient Outcome(s) Required Intervention;
Patient Age81 YR
-
-