Model Number 989605414121 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Return of the suspect transducer is anticipated.Evaluation results will be included in a follow-up report upon its return and investigation completion.
|
|
Event Description
|
A customer reported intermittent articulation issues with their x7-2t model transducer.There was no patient injury associated with this event.
|
|
Manufacturer Narrative
|
Evaluation of the transducer could not duplicate the articulation issue as described by the customer.Investigation of the device noted a hole in the bending neck sheath, scuff mark on the lens and debris on the handle switch.Additional functional testing of the device confirmed the mechanical movement of this device was functioning as designed and no articulation impediment was identified.
|
|