MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VTICMO12.6
Device Problems Positioning Failure (1158); Entrapment of Device (1212); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/28/2017
Event Type  malfunction  
Manufacturer Narrative
This product is not marketed in the us.(b)(4).
 
Event Description
The reporter indicated that the surgeon attempted to implant a 12.6 mm vticmo12.6, -12.50/2.5/117 (sphere/cylinder/axis), implantable collamer lens in the patient's left eye (os) on (b)(6) 2017.The lens was "half implanted" when the plunger overrode the lens and got stuck in the cartridge.When the lens was removed for reloading, a tear in the peripherical haptic was noted.The lens was not reimplanted.On (b)(6) 2017, a similar lens was implanted and it was successful.
 
Manufacturer Narrative
Additional data: device evaluation: the lens was returned in liquid inside a lens vial.Visual inspection found no visible damage to the lens.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
MDR Report Key7161181
MDR Text Key96355900
Report Number2023826-2018-00001
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor,foreign
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date09/30/2020
Device Model NumberVTICMO12.6
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/05/2017
Initial Date FDA Received01/03/2018
Supplement Dates Manufacturer Received03/20/2018
Supplement Dates FDA Received03/23/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age34 YR
-
-