STRYKER ENDOSCOPY-SAN JOSE PKG, 5MM PEEK MONOPOLAR HANDLE; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED Back to Search Results
Catalog Number 0250080617
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/09/2017
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.(b)(4).
 
Event Description
It was reported that the insulation had been compromised.
 
Event Description
It was reported that the insulation had been compromised.
 
Manufacturer Narrative
(b)(4).Alleged failure: burn on insulation at shaft.Confirmed failure: burn on insulation at shaft.Probable root cause: poor autoclave reliability, incorrect sterilization/reprocessing procedure, handling procedures, contact forces, product used beyond defined useful life.The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, 5MM PEEK MONOPOLAR HANDLE
Type of Device
ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key7165608
MDR Text Key96495182
Report Number0002936485-2018-00022
Device Sequence Number1
Product Code OCZ
Combination Product (y/n)N
PMA/PMN Number
K973259
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 04/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0250080617
Device Lot Number1445328H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2017
Initial Date Manufacturer Received 12/09/2017
Initial Date FDA Received01/04/2018
Supplement Dates Manufacturer Received12/09/2017
Supplement Dates FDA Received04/06/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-