• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, ROLLER, 24-28 FR., 12° AND 30°, STERILE, SINGLE USE, 12; HF-RESECTION ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, ROLLER, 24-28 FR., 12° AND 30°, STERILE, SINGLE USE, 12; HF-RESECTION ELECTRODES Back to Search Results
Model Number WA22351C
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Abdominal Pain (1685); Bowel Burn (1756); Fever (1858); Internal Organ Perforation (1987); Peritonitis (2252); Constipation (3274)
Event Type  Injury  
Manufacturer Narrative
The suspect medical device was not returned to olympus for evaluation/investigation since it was reportedly discarded by the user facility.Therefore the exact cause of the patient's outcome and the reported phenomenon could not be determined and is being judged as unknown.Furthermore, a manufacturing and quality control review could not be performed since basic data of article identification (lot number) are missing.The case will be closed from olympus side with no further actions.However, the reported phenomenon will be recorded for trending and surveillance purposes.
 
Event Description
Olympus was informed that some days after a transurethral resection of the bladder tumor in saline (turis-bt) procedure, the patient returned to the hospital with fever, abdominal pain and constipation.A ct scan revealed a peritonitis and a perforation of the small intestine was found and treated during a follow-up laparotomy.However, there were no reports of any device malfunctions and the turis-bt procedure was completed with the same set of equipment.Furthermore, the patient reportedly recovered after the laparotomy.
 
Manufacturer Narrative
A material or quality problem can be excluded since a manufacturing and quality control review was performed for the eligible lot numbers of the hf resection electrode without showing any abnormalities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, ROLLER, 24-28 FR., 12° AND 30°, STERILE, SINGLE USE, 12
Type of Device
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer (Section G)
OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O.
telickova 457/29
prerov
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 
4940669662
MDR Report Key7166970
MDR Text Key96441957
Report Number9610773-2018-00002
Device Sequence Number1
Product Code FAS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK100275
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberWA22351C
Device Catalogue NumberWA22351C
Other Device ID Number14042761051689
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/14/2017
Initial Date FDA Received01/05/2018
Supplement Dates Manufacturer Received02/02/2018
Supplement Dates FDA Received02/02/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-