ORTHO CLINICAL DIAGNOSTICS REAGENT RED BLOOD CELLS 0.8% RESOLVE PANEL A Back to Search Results
Catalog Number 6902317
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/15/2017
Event Type  malfunction  
Manufacturer Narrative
Ortho performed retain testing, batch review, complaint review by lot, donor history, and donor complaint review.Based upon the results of this investigation, the reported customer issue was unable to be confirmed.All results were satisfactory.Sample was not returned to ortho for further investigation.(b)(4).
 
Event Description
Event 3 of 4 (patient 2 tested with vision).Customer contacted tsc to report two patients with history of anti-e failed to react with cell# 3 from 0.8% resolve panel a lot# vra290 exp: jan 2, 2018; patients tested with mts anti-igg gel cards on vision and manual gel.Customer reports patients in question, antibody screen initially tested 1+ positive with cell# 2 from 0.8% selectogen on vision with same lot of mts igg gel cards.Customer reports routine qc performed on 0.8% resolve panel a in question not affected.Customer reports no patient harm caused due to this event occuring.Issue started on: (b)(6) 2017.Frequency: two patients both manual and on vision.Microtubes/wells or cell (donor #) affected: cell#3.Methodology used: vision.Reaction grade obtained: neg.Customer was expecting: pos.Test repeated: yes.Result obtained by repeating: neg.Method used to repeat: manual gel.Customer reports no signs of hemolysis in red cell reagents.Customer reports storing and testing mts gel cards and reagents according to ifu.Customer reports performing antibody identification with 0.8% resolve panel b and all homozygous b cells reacted 1+ pos with patients in question.Customer reports if they did not have 0.8% resolve panel b, they would have failed to identify anti-e in patients in question.Detail any maintenance failure or maintenance inadequately performed that would be relevant for the issue : up to date.Tsc discussed with customer to phenotype e antigen on cell#3 from 0.8% resolve panel a lot# vra290 in question.Customer agreed and reports 4+ positive reaction for e antigen in tube method (lot numbers requested and not provided).Tsc discussed with customer that patients in question may have been expressing anti-e at sub-detectable levels.Tsc referred and discussed ifu with customer: for antibody detection and identification, different serological methods are optimal for different antibodies and that no single antibody screening or identification method optimally detects all antibodies.In some low ionic strength test systems, certain anti-e and anti-k antibodies have been reported to be nonreactive.Customer content with discussion and requests replacements for the one set of 0.8% resolve panel a.Tsc will process replacement and email customer certificate of destruction, customer agreed.Customer content with documentation of event and replacement and agreed to call back if further assistance needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAGENT RED BLOOD CELLS 0.8% RESOLVE PANEL A
Type of Device
REAGENT RED BLOOD CELLS
Manufacturer (Section D)
ORTHO CLINICAL DIAGNOSTICS
1001 route 202
raritan NJ 08869
Manufacturer (Section G)
ORTHO CLINICAL DIAGNOSTICS
1001 route 202
raritan NJ 08869
Manufacturer Contact
matthew p wictome
1001 route 202
raritan, NJ 08869
9082188223
MDR Report Key7169355
MDR Text Key97738911
Report Number2250051-2018-00005
Device Sequence Number1
Product Code KSZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Medical Technologist
Type of Report Initial
Report Date 01/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Medical Technologist
Device Expiration Date01/02/2018
Device Catalogue Number6902317
Device Lot NumberVRA290
Other Device ID Number10758750007677
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/17/2017
Initial Date FDA Received01/05/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/31/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-