Model Number SLIMLINE SIS 200¿ FIBER
Device Problem Material Fragmentation (1261)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/08/2017
Event Type  malfunction  
Manufacturer Narrative
An investigation of the reported event found that the reported malfunction of the slimline 200 fiber was similar to a device malfunction that was alleged to have caused or contributed to a potentially harmful situation.Because the company is aware that this malfunction was alleged to have caused or continued to a potentially harmful situation (reference mdr #3004135191-2017-00002) this event represents a reportable malfunction.The complainant indicated that the device has been disposed and will not be returned for evaluation; therefore a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event Description
A user facility reported that while a physician was using a slimline 200 laser fiber during a bladder stone procedure, the tip of the fiber broke off in the patient.The tip was washed out of the bladder with cystoscope, and the procedure was completed with another slimline 200 fiber.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLIMLINE SIS 200¿ FIBER
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
LUMENIS LTD.
6 hakidma street
po box 240
yokneam, 20692
IS  20692
Manufacturer (Section G)
LUMENIS LTD.
6 hakidma street
po box 240
20692
IS   20692
Manufacturer Contact
alan vaisman
6 hakidma street
po box 240
IS   20692
MDR Report Key7170838
MDR Text Key97641653
Report Number3004135191-2018-00009
Device Sequence Number1
Product Code GEX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140388
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 01/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date10/09/2021
Device Model NumberSLIMLINE SIS 200¿ FIBER
Device Catalogue Number0644-010-01
Device Lot Number73991116
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/23/2017
Initial Date FDA Received01/07/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/09/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-