ARTHROCARE CORP. SMARTSTITCH PP CONNECTOR; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Catalog Number OM-8010
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/04/2018
Event Type  malfunction  
Event Description
It was reported that the jaw broke off at the joint.The broken piece was removed and the procedure was completed using a back-up device.No patient injury or other complications were reported.
 
Manufacturer Narrative
The reported perfectpasser connector, intended for use in treatment, was returned for evaluation.A relationship between the device and reported incident was established.From the information provided, the jaw broke off at the joint.Visual inspection shows the connector was received with its s-clamp unhooked from the s-hook.The clamp was not broken as indicated by the customer; however, it did come loose.Internal investigation indicates the failure cause for upper s-clamp unhooked is due to a dimensional discrepancy on the s-hook component.Changes to the component have been implemented to prevent this failure.The instruction for use (ifu) was reviewed and was found to outline precautionary statements and instructions in regards to the use of the device to avoid damage or non-functionality.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.Device returned for evaluation.Device evaluated by the manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTSTITCH PP CONNECTOR
Type of Device
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
jim gonzales
7000 w. william cannon
austin, TX 78735
MDR Report Key7173551
MDR Text Key96813801
Report Number3006524618-2018-00023
Device Sequence Number1
Product Code NBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123268
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberOM-8010
Device Lot Number1194208
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/04/2018
Initial Date FDA Received01/08/2018
Supplement Dates Manufacturer Received01/29/2018
Supplement Dates FDA Received02/01/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-