Model Number MMT-399T
Device Problem Bent (1059)
Patient Problem Hyperglycemia (1905)
Event Date 07/30/2016
Event Type  Injury  
Manufacturer Narrative
Note: this is a remediation mdr.Medtronic diabetes implemented revised mdr reportability criteria effective on october 1, 2017.Subsequently, medtronic diabetes conducted a two year retrospective review of complaints.This event was retrospectively identified to be reportable based on the revised mdr reportability criteria.Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event Description
Customer reported via phone call that they experienced a high blood glucose level due to a bent infusion set cannula.Customer's blood glucose level was 444 mg/dl.Customer contacted their doctor after which they treated the high blood glucose level with a manual injection of insulin using a syringe.Customer was able to troubleshoot.Customer's reported that the infusion set cannula was bent at a 90 degree angle upon removal.Customer did not troubleshoot the elevated blood glucose level.Customer was able to troubleshoot the bent infusion set cannula.No product is expected to be returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM QUICK SET MMT-399T
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire st.
northridge, CA 91325-1219
MDR Report Key7175487
MDR Text Key96749888
Report Number2032227-2018-00506
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 01/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-399T
Device Catalogue NumberMMT-399T
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/30/2016
Initial Date FDA Received01/09/2018
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age65 YR
Patient Weight70
-
-