• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC ACE PLUS 7 W ADV HEMOSTASIS; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC ACE PLUS 7 W ADV HEMOSTASIS; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HARH36
Device Problems Break (1069); Loose or Intermittent Connection (1371)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/15/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # p92n9h.The lot history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
 
Event Description
It was reported that the ceramic of the jaw of the device broke loose.The ceramic was removed from the patient.Another product was opened to continue the surgery.There was no patient harm, surgery was not prolonged.
 
Manufacturer Narrative
(b)(4).Date sent: 04/23/2018.Batch # p92n9h.Device analysis: the analysis results found that the device was received with the tissue pad detached and not returned but with evidence of body fluids and tissue pad material in the groove section of the clamp arm.In addition with the blade tip intact and not broken off as reported by the customer.The device was connected to a test hand piece and a gen11 and the device did activate during functional testing.The device was disassembled to inspect the internal components and no anomalies were found.Based on the condition of the tissue pad, a probable cause of this damage is that the clamp of the device may have been closed and the instrument activated without tissue present.Care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad.The resulting damage contributes to the removal of the pad from the clamp arm.The cleaning of the pad, not in accordance with the ifu, can also result in removal of the pad during use.It is possible that the device returned may have been the one used to complete the procedure and it is not the device complained about.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC ACE PLUS 7 W ADV HEMOSTASIS
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key7178546
MDR Text Key96950184
Report Number3005075853-2018-07184
Device Sequence Number1
Product Code LFL
UDI-Device Identifier20705036014628
UDI-Public20705036014628
Combination Product (y/n)N
PMA/PMN Number
K132612
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 12/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2022
Device Catalogue NumberHARH36
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2018
Initial Date Manufacturer Received 12/15/2017
Initial Date FDA Received01/10/2018
Supplement Dates Manufacturer Received03/26/2018
Supplement Dates FDA Received04/23/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-