Catalog Number HP054
Device Problems Loose or Intermittent Connection (1371); Difficult to Remove (1528)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/01/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # p9270k.Investigation summary the handpiece was received with the nose cone cracked and the mount was noted to be loose.No functional testing could be performed due to the nose cone was extremely cracked and no instrument could be attached to the handpiece.The instrument was disassembled to inspect internal components.The moisture indicator was positive.The transducer assembly was not held in place due to the cracked nose cone, so torquing on the disposable resulted in twisting only the handpiece transducer assembly until the internal wires got disconnected.Due to the cracked nose cone, moisture entered the hand piece mid housing.Analysis was unable to determine the exact root cause that lead to crack in the hand piece nosecone.It is possible that the ingress of moisture affected handpiece functionality.The batch history record was reviewed and no defects, nc¿s or protocols related to the complaint, were found during the manufacturing process.
 
Event Description
It was reported that prior to a gastric sleeve procedure the scrub tech was unable to tighten the instrument to the hand piece.There were no clicks, it just spun.They were also unable to remove the instrument from the hand piece.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC SCALPEL HAND PIECE
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7179230
MDR Text Key96941279
Report Number3005075853-2018-07201
Device Sequence Number1
Product Code LFL
UDI-Device Identifier20705036002144
UDI-Public20705036002144
Combination Product (y/n)N
PMA/PMN Number
K002906
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial
Report Date 12/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberHP054
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2017
Initial Date Manufacturer Received 12/18/2017
Initial Date FDA Received01/10/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/04/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-