A siemens customer service engineer (cse) was sent to the customer's site for system inspection.The cse checked the air pump calibrations, sample probe diluter fluidics, and adjustments were made to the air pump flow range.The cse checked quality control, and a patient run that were acceptable.The cause of the non-reproduced false low advia centaur cp total human chorionic gonadotropin (hcg) result may be due to sampling issues.No additional concerns have been reported by the customer.The instruction for use (ifu) under the interpretation of results section states the following: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings." the instruction for use (ifu) under the limitation section states the following: "all in vitro assays can generate erroneous results, both clinically false positive results (test results suggesting a condition that is absent) and clinically false negative results (test results failing to identify a condition that is present)." "there are many possible causes for these types of discordant results.Erroneous results may occur due to interference from identifiable serum constituents or patient-specific serum constituents." "if an aberrant or abnormal result, as defined by the laboratory protocol, occurs, laboratory personnel should first make certain that the system is performing and is operated and maintained in accordance with the product labeling.The user should then follow the laboratory protocol for advising the clinician of a result that appears to have deviated from the norms established by the laboratory." "test results alone are not diagnoses of medical conditions.For example, low titer elevations of hcg can occur in normal non-pregnant subjects.A physician's diagnosis involves evaluation of the test result in conjunction with, and in the context of, the patient's medical history, physical examination, and other test results sometimes in consultation with other medical experts." the instrument is performing within specification.No further investigation is required.
|
A false low advia centaur cp total human chorionic gonadotropin (hcg) result obtained by the customer on a patient sample, and the reported result was questioned by the physician (patient).The same sample was repeated the following day, and the result was higher.The patient was redrawn, tested on a different analyzer, and the result was high.A corrected report was issued.There was no report of patient treatment being prescribed or altered, or adverse health consequences due to the discordant advia centaur cp total hcg result.
|