• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Model Number N/A
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/22/2017
Event Type  malfunction  
Manufacturer Narrative
This complaint is recorded with zimmer biomet under (b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, product evaluated by external contractor.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the unit was leaking fluid.The event timing was unknown.No adverse events have been reported as a result of the malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The device history record (dhr) review for intellicart system serial number (b)(4), noted no related non-conformances, requests for deviation (rfd), change notices (cn), or any other issues with manufacturing.The dhr review found no issues with the device and all verifications, inspections, and tests were successfully completed.The device was noted to have not been previously repaired by zimmer biomet surgical.On (b)(6) 2017, it was reported from (b)(6) that the unit was leaking fluid.Pm medical was contacted about the unit and dispatched a service technician to be at the site.On (b)(6) 2017, the technician found that the unit was issuing drain error and performed the backflush to resolve the drain error.While performing the backflush, he noted that the unit was leaking fluid from the receiver and suspected the leaking was due to back pressure.He performed a full soak cycle and then verified that the cart was functioning as intended.The technician then returned the cart to service without further incident.While the service technician confirmed the reported event and the device was noted to be functioning as intended after a full soak was performed, it is unknown with the information provided as to what actually caused the leaking from cart.Therefore, based on the information provided, a specific root cause of the reported event cannot be determined.The investigation was based on the information that was provided initially and any information that was.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO FLUID CART WITH SMOKE EVACUATOR
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer (Section G)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key7186086
MDR Text Key97399668
Report Number0001954182-2018-00004
Device Sequence Number1
Product Code JCX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK162421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00514010200
Device Lot Number0025013
Other Device ID Number(01) 00889024466005
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/22/2017
Initial Date FDA Received01/12/2018
Supplement Dates Manufacturer Received01/19/2018
Supplement Dates FDA Received01/19/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/24/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-