• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number MICL13.2
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/06/2017
Event Type  malfunction  
Manufacturer Narrative
Device evaluation: product evaluation found that the lens was returned in a specimen cup and an empty cartridge in the cup.Visual inspection found no visible damage to lens.Work order search: no similar complaint type events were reported for units within the same lot.(b)(4).
 
Event Description
The reporter indicated that the surgeon noticed microscopic black flakes on a micl13.2, -4.0 diopter, implantable collamer lens when they opened the lens vial and were preparing to load the lens in the cartridge for the injection on (b)(6) 2017.The surgeon described opening the lens vial with hemostats, removing the metal ring, and removing the grey stopper with a sterile gloved hand.The lens was poured out of the vial onto the injector cartridge tray with the assistance of the injector plunger to guide the lens out of the vial.When the lens was picked up with icl loading forceps, the black flakes were noticed.There was no patient contact.They used a new cartridge/injector and a new icl was loaded and used to perform the surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
MDR Report Key7188510
MDR Text Key97750078
Report Number2023826-2018-00067
Device Sequence Number1
Product Code MTA
UDI-Device Identifier00841542103176
UDI-Public00841542103176
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 12/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date10/31/2018
Device Model NumberMICL13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/18/2017
Initial Date FDA Received01/12/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/11/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-