• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL, INC 3100 HIGH FREQUENCY OSCILLATORY VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL, INC 3100 HIGH FREQUENCY OSCILLATORY VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100B
Device Problems Thermal Decomposition of Device (1071); Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/20/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).At this time, vyaire medical has not received the suspect device/component from the customer for evaluation.In the event that the device is received for evaluation or additional information is received, a follow-up report will be submitted.
 
Event Description
The customer reported that the ventilator stopped, while in use on a patient.The customer noted a burning smell right before the piston stopped.The patient was placed on another ventilator and there was no patient compromise associated with this reported issue.
 
Manufacturer Narrative
Failure analysis results: vyaire medical was not able to duplicate the customer's reported issue during the failure analysis process.The power driver module was received without any physical damage to it.After installing the pcba into a known good top level unit, the service technician was able to perform the circuit calibration and the system was operational.The system cycled for 5 hours and no performance issues were found.The power was then cycled on/off a total of 10 times with no abnormalities.There were no performance issues found.The device passed all testing and met all manufacturer specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATORY VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
VYAIRE MEDICAL, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
kristin graf
22745 savi ranch parkway
yorba linda, CA 92887
MDR Report Key7189913
MDR Text Key97279853
Report Number2021710-2018-07299
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100B
Device Catalogue Number770155
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/20/2017
Initial Date FDA Received01/15/2018
Supplement Dates Manufacturer Received12/20/2018
Supplement Dates FDA Received01/25/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-