Model Number 3100B |
Device Problems
Thermal Decomposition of Device (1071); Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/20/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).At this time, vyaire medical has not received the suspect device/component from the customer for evaluation.In the event that the device is received for evaluation or additional information is received, a follow-up report will be submitted.
|
|
Event Description
|
The customer reported that the ventilator stopped, while in use on a patient.The customer noted a burning smell right before the piston stopped.The patient was placed on another ventilator and there was no patient compromise associated with this reported issue.
|
|
Manufacturer Narrative
|
Failure analysis results: vyaire medical was not able to duplicate the customer's reported issue during the failure analysis process.The power driver module was received without any physical damage to it.After installing the pcba into a known good top level unit, the service technician was able to perform the circuit calibration and the system was operational.The system cycled for 5 hours and no performance issues were found.The power was then cycled on/off a total of 10 times with no abnormalities.There were no performance issues found.The device passed all testing and met all manufacturer specifications.
|
|