Catalog Number 0250181106
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/19/2017
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.(b)(4).
 
Event Description
It was reported that the insulation had been compromised.
 
Manufacturer Narrative
Alleged failure: needs re-insulated.The probable root causes could be user misuse, excessive force, normal wear, or improper sterilization methods.Although the product passed the insulscan, the failure identified in the investigation is consistent with the complaint record because there were dents and scratches on the insulation.The product was returned for investigation and the failure mode was confirmed and will be monitored for future reoccurrence.Gtin: (b)(4).
 
Event Description
It was reported that the insulation had been compromised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, 5MM MULTI-FUNCTION PEEK HANDLE
Type of Device
ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
ashley lower
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key7191106
MDR Text Key97480956
Report Number0002936485-2018-00058
Device Sequence Number1
Product Code OCZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973259
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0250181106
Device Lot Number1542101H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/19/2017
Initial Date FDA Received01/15/2018
Supplement Dates Manufacturer Received12/19/2017
Supplement Dates FDA Received04/12/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-