• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VTICH13.2
Device Problems Improper or Incorrect Procedure or Method (2017); Device Dislodged or Dislocated (2923)
Patient Problem No Code Available (3191)
Event Date 11/30/2017
Event Type  Injury  
Manufacturer Narrative
This product is not marketed in the us.(b)(4).
 
Event Description
The reporter indicated that the surgeon implanted a 13.2 mm vtich13.2, 7.50/4.5/105 (sphere/cylinder/axis), implantable collamer lens in the patient's left eye (os) on (b)(6) 2017.On (b)(6) 2018, the lens was explanted due to the haptic being torn during implantation and lens rotation.The lens was exchanged for a similar lens and the problem was resolved.It was reported that the lens tear was due to the surgeon not using enough hpmc.
 
Manufacturer Narrative
Device evaluation: product evaluation found the lens returned dry in a micro centrifuge vial with clear residue on lens.Visual inspection found pieces of the haptic missing and residue on lens.A lens work order search was performed.No similar complaint type events were found.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
MDR Report Key7197912
MDR Text Key97469651
Report Number2023826-2018-00074
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date09/30/2020
Device Model NumberVTICH13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/19/2017
Initial Date FDA Received01/16/2018
Supplement Dates Manufacturer Received03/13/2018
Supplement Dates FDA Received03/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age28 YR
-
-