Model Number SC-2218-50
Device Problems Migration or Expulsion of Device (1395); Material Separation (1562)
Patient Problem Tissue Breakdown (2681)
Event Date 10/13/2017
Event Type  Injury  
Manufacturer Narrative
Additional suspect device component involved in the event: model#: sc-2218-50, serial#: (b)(4).Description: linear st lead, 50cm the explanted device was not returned to bsn.
 
Event Description
A report was received that patient¿s lead had migrated significantly, the physician went through it and found out that there were sutures broke through tissues which caused the clik anchor to dislodge and this was confirmed through x-ray.It was also reported that there were no sutures and no abandoned clik anchor that was left in the patient¿s body.No device malfunction was suspected.The patient underwent a lead replacement procedure and was doing well postoperatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION SPECTRA
Type of Device
SPINAL CORD STIMULATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key7201525
MDR Text Key97591865
Report Number3006630150-2017-05457
Device Sequence Number1
Product Code LGW
UDI-Device Identifier08714729767725
UDI-Public08714729767725
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date11/07/2018
Device Model NumberSC-2218-50
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/19/2017
Initial Date FDA Received01/17/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/10/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age59 YR
-
-