Model Number U9525
Device Problem Cut In Material (2454)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Upon triage the service tech found the pump to have a damaged power cord with exposed copper on the base of the cord.
 
Manufacturer Narrative
The scd express controller was received with a complaint of ¿exposed copper on base of power cord¿.The complaint was verified as the power cord shows bare copper wire near the base of the unit.A physical analysis was performed on the unit and there was no gross damage noted.The damaged cord was removed and retained.The power cord was replaced per factory service procedures.The potential root cause is due to the procedure used to unplug the unit and the procedure of wrapping of the cord around the bed hook.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KENDALL SCD
Type of Device
SLEEVE, LIMB, COMPRESSIBLE
Manufacturer (Section D)
COVIDIEN MFG PLAINFIELD
2824 airwest blvd
plainfield IN 46168
Manufacturer (Section G)
COVIDIEN MFG PLAINFIELD
2824 airwest blvd
plainfield IN 46168
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key7203859
MDR Text Key97904278
Report Number3008361498-2018-00034
Device Sequence Number1
Product Code JOW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberU9525
Device Catalogue NumberU9525
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/17/2018
Initial Date FDA Received01/18/2018
Supplement Dates Manufacturer Received01/17/2018
Supplement Dates FDA Received02/06/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-