Model Number MICL13.2
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
No similar complaint type events were reported for units within the same lot.(b)(4).
 
Event Description
The reporter indicated that the surgeon noticed a microscopic black speck on a micl13.2, -9.5 diopter, implantable collamer lens.The lens was poured out of the vial into the packaging of the cartridge and that's when the speck was noticed.It was reported that the icl had touched nothing but the plunger which was directly opened and used to slide the icl out.The speck was located on the surface of the lens.
 
Manufacturer Narrative
Device evaluation: the lens was returned in liquid, in a lens case/vial.Visual inspection found no visible damage to the lens.The lens was inspected under 20x on the microscope and no black spec was noted.Claim # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
MDR Report Key7208090
MDR Text Key97950210
Report Number2023826-2018-00133
Device Sequence Number1
Product Code MTA
UDI-Device Identifier00841542103282
UDI-Public00841542103282
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date04/30/2018
Device Model NumberMICL13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/20/2017
Initial Date FDA Received01/19/2018
Supplement Dates Manufacturer Received07/03/2018
Supplement Dates FDA Received08/02/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/17/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-