• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTECH SA ROSA SURGICAL DEVICE; COMPUTER-ASSISTED SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTECH SA ROSA SURGICAL DEVICE; COMPUTER-ASSISTED SURGICAL DEVICE Back to Search Results
Model Number ROSA 2.5.8
Device Problems Bent (1059); Erratic or Intermittent Display (1182)
Patient Problem No Patient Involvement (2645)
Event Date 01/16/2018
Event Type  malfunction  
Manufacturer Narrative
The device has not been evaluated yet for investigation purpose.Once the evaluation will be performed, a follow-up medwatch report will be submitted.
 
Event Description
It was reported that the power input connector is bent due to long-term-use effort applied via the power cable.The monitor is still functional, but depending on the position of the touchscreen, there are some powering off/up issues.
 
Manufacturer Narrative
Following the investigation, the root cause of the bent of the power input connector is wear due to long-term-use effort applied via the power cable.Corrected data: date of this report, date received by manufacturer, if follow-up, what type, device evaluated by manufacturer, evaluation code, remedial action.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA SURGICAL DEVICE
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
languedoc-roussillon 34000
FR  34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
Manufacturer Contact
jay sharma
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key7214628
MDR Text Key98046499
Report Number3009185973-2018-00024
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
PK101791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Remedial Action Replace
Type of Report Initial,Followup
Report Date 02/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Medical Equipment Company Technician/Representative
Device Model NumberROSA 2.5.8
Device Catalogue NumberN/A
Device Lot NumberN/A
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/16/2018
Initial Date FDA Received01/24/2018
Supplement Dates Manufacturer Received02/01/2018
Supplement Dates FDA Received02/01/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/11/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction NumberN/A
Patient Sequence Number1
-
-