ST. JUDE MEDICAL, INC. EP-WORKMATE¿ SCU-SIGNAL CONDITIONING UNIT (AMPLIFIER), 120 CHANNEL; COMPUTER, DIAGNOSTIC, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 09-1463-0120 |
Device Problem
Noise, Audible (3273)
|
Patient Problem
No Information (3190)
|
Event Date 01/04/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
The reported incident could not be reproduced during the hardware evaluation.The results of the hardware evaluation confirmed proper functions of the returned amplifier such as communication, signals acquisition/amplification and data processing.No abnormal operating symptoms or unexpected abnormal functions under normal operating condition were observed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with manufacturing specifications and procedures.Based on the information provided and the hardware evaluation performed, the cause of the reported noise could not be conclusively determined.
|
|
Event Description
|
During the procedure, noise was noted when selecting the pacing channel without pacing.Troubleshooting included checking settings, changing the ecg cable and pin box, and changing the location on the pin box.The procedure was cancelled.
|
|