(b)(4).Batch # p5552z.Device evaluation: the analysis results found that the ats45 device was received with the firing mechanism damaged and with 6r45b cartridge loaded in the device.The reload was received partially fired 1/10 and with the lockout spring normal.No functional test could be performed due to the condition of the device.The device was disassembled to verify the condition of the internal components and the firing trigger teeth were found broken.While no conclusion could be reached on what caused the firing mechanism to fail, it is possible that the device was attempted to fire on ticker tissue then indicated or attempted to fire trough a locked reload in previous firings causing an increase of the internal forces resulting in the component yielding.It should be noted that a 100% inspections takes place during manufacturing to ensure the device meets the require specifications; in addition, a sample of the batch is inspected at fgqa.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.Attempts have been made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.When the device did not perform stapling, was the tissue cut (tissue cut and no staples in cut area)? was there another defect with the device? if yes, please explain.Was there any patient consequence or change in the post-operative care of the patient as a result of the event?.
|