• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ETS FLEX ARTICNG LNR CUTR 45MM; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ETS FLEX ARTICNG LNR CUTR 45MM; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number ATS45
Device Problems Defective Component (2292); Failure to Form Staple (2579)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/23/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # p5552z.Device evaluation: the analysis results found that the ats45 device was received with the firing mechanism damaged and with 6r45b cartridge loaded in the device.The reload was received partially fired 1/10 and with the lockout spring normal.No functional test could be performed due to the condition of the device.The device was disassembled to verify the condition of the internal components and the firing trigger teeth were found broken.While no conclusion could be reached on what caused the firing mechanism to fail, it is possible that the device was attempted to fire on ticker tissue then indicated or attempted to fire trough a locked reload in previous firings causing an increase of the internal forces resulting in the component yielding.It should be noted that a 100% inspections takes place during manufacturing to ensure the device meets the require specifications; in addition, a sample of the batch is inspected at fgqa.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.Attempts have been made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.When the device did not perform stapling, was the tissue cut (tissue cut and no staples in cut area)? was there another defect with the device? if yes, please explain.Was there any patient consequence or change in the post-operative care of the patient as a result of the event?.
 
Event Description
It was reported that during an appendectomy the device was defective and did not perform stapling during surgery.The device was replaced by another.There were no patient consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETS FLEX ARTICNG LNR CUTR 45MM
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7222276
MDR Text Key98859670
Report Number3005075853-2018-07560
Device Sequence Number1
Product Code GDW
UDI-Device Identifier20705036001079
UDI-Public20705036001079
Combination Product (y/n)N
PMA/PMN Number
K020779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 11/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2021
Device Catalogue NumberATS45
Device Lot NumberP4R27W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2018
Initial Date Manufacturer Received 01/08/2018
Initial Date FDA Received01/26/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/19/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-